Description
1]P.O.: Power switch (green lamp): If the oxygen machine is connected to electricity it will be green
2]P.F.: Power failure(red lamp): It means a power failure alarm, when no electricity, the fuse is damaged, the red light will turn on and an audible alarm meanwhile.
3]L.P.: low pressure(yellow lamp): Low-Pressure alarm, means When the inside system pressure is below 0.12, the light will be red and you may hear a continuous alarm, meaning there is some air leakage in the machine.
4]It is mandatory to use a voltage stabilizer. Make sure to use 1 KV stabilizer as a must to maintain the life of the Oxygen Concentrator. We do not recommend the inverter, as the machine performance might get affected, make sure to use 1 KV stabilizer if you happen to use it. Always have a backup cylinder on hand.
Timing buttons:
The two buttons are used for timing adjustment, and each press of the left button(△) will increase timing by 10min, the max timing is 40 hours. And each press of the right button(▽) will decrease timing by 10min. When the right button(▽) is pressed to reduce timing to “0”, the medical oxygen concentrator will turn off automatically.
Humidifier:
Humidifier which is used for humidifying oxygen and preventing throat and nasal mucosa stimulated by dry oxygen and dry hard sputum difficult to spit out
1]Model: GVS 5L Oxygen Dual Flow
2]Flow Rate: 1-5 LPM
3]Purity: 93%(±3%)
4]Outlet Pressure: 0.04-0.07Mpa\
6]Sound Level: ≤43db
7]LCD Display: Accumulating timing, Present timing, Timing
8]Alarm: Power failure alarm, high/low-pressure alarm
10]Standard Accessories: Oxygen Concentrator, Oxygen tube, Humidifier bottle, Primary air suction filtering sponge, Secondary air suction filters, General Machine Manual(SPO2 port is optional which is not included in this machine )
Warranty:
The product from the date of sale, under normal use and storage conditions, the warranty period of this product is 24 Months or 5000 Hrs